II) V24/RS232/485 Interface - Single Analyzers and Systems

AKAK – Read command "Calibration gas concentration"

To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device) and range:

The calibration gas concentration in ppm.

Read command

AKAK Kn [Mx]

Range number (optional); x = 1, 2, 3, 4 Output of all ranges, if no range is named.

Read of channel n

Code

Response

AKAK 0 M1 XXX M2 XXX ... Mn ZZZ

Concentration of calibration gas

Range n

Concentration of calibration gas

Range 2

Concentration of calibration gas

Range 1

Error status

Code

The values will get the same format for the read of single ranges.

90003752(1) [AK-Commands] 10/98

AK

2 - 53