AKAL – Read command "Stored calibration corrections"

To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device):

The corrections in ppm determined and stored during the last calibration. These corrections are also necessary to calculate the analyzer values (deviations from the linearized curve).

The sum of deviations (total correction) to the calibration before the last linearization check for the last range calibrated with zero and spangas.

Read command

AKAL K0 [Mx]

AKAL Kn [Mx]

Range number (optional); x = 1, 2, 3, 4 Output of all ranges, if no range is named.

Read of channel n

Code

2 - 54

AK

90003752(1) [AK-Commands] 10/98