Instruction Sheet

PN 51A-404/rev.M December 2010

Model 404

ENDURANCEConductivity Sensor

For additional information, please visit our website

at www.emersonprocess.com/raihome/liquid/.

CAUTION

SENSOR/PROCESS

APPLICATION COMPATIBILITY

The wetted sensor materials may not be compatible with process composition and operating conditions. Application compatibility is entirely the responsi- bility of the user.

WARNING

Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!

SENSOR SPECIFICATIONS

SPECIFICATIONS

MODEL 404-16 (PVC body)

MODEL 404-17 (stainless steel body)

 

 

 

Wetted Materials

Titanium, PEEK (glass filled), PVC,

Titanium, PEEK (glass filled),

 

EPDM, polyethylene

303 SST, EPDM

 

 

 

Temperature Range

32-140°F (0-60°C)

32-212°F (0-100°C)

 

 

 

Maximum Pressure

20 psig (239 kPa abs) at 60°C (140°F)

100 psig (791 kPa abs)

 

100 psig (791 kPa abs) at 25°C (77°F)

 

 

(use FNPT fittings only)

 

 

 

 

INSTALLATION

If the sensor is installed in a sidestream with the sample draining to open atmos- phere, bubbles may accumulate on the electrodes. Trapped bubbles will cause errors. Normally, as bubbles accumu- late the conductivity reading drifts down. To control bubble formation, apply a small amount of back pressure to the sensor.

FIGURE 1. Installation of Model 404 Sensor

Page 1
Image 1
Emerson PN 51A-404 instruction sheet Sensor Specifications, Installation, SENSOR/PROCESS Application Compatibility