Installation, Operation & Maintenance Manual
IP2040/IM, Rev. AA
MSP900SH |
|
|
|
|
| May 2007 |
|
|
|
|
|
|
|
2.2 Safety Data |
|
|
|
|
|
|
Type numbers |
|
| ||||
Certificate numbers | ATEX | BAS01ATEX1061X | FM | 3021193 | ||
ATEX Coding | CSA | 1352094X | CSA | 1352094X | ||
|
|
|
|
|
| |
(EU Directive 94/9/EC) |
| II I G |
|
|
|
|
Cenelec Coding |
| EEx ia IIC T4 |
|
|
| |
|
| EEx ia IIC T6 |
|
|
| |
CSA Coding |
| Ex ia IIC T4 | Ex ia IIC T4 | |||
|
| Ex ia IIC T6 | Ex ia IIC T6 | |||
FM Coding |
|
|
| IS/I/1/ABCD T6 Ta = 55°C | ||
|
|
|
|
|
| T4 Ta = 60°C |
|
|
|
| IS/I/0/AEx ia IIC T6 Ta = 55°C | ||
|
|
|
|
|
| T4 Ta = 60°C |
Safety parameters |
| Ui | 30 V | Ui (Vmax) | 30 V | |
|
| |||||
|
| Ii | 120mA | Ii (lmax) | 120mA | |
|
| Pi | 0.82W | Pi |
| 0.82W |
|
| Li | 27mH | Li |
| 27mH |
|
| Ci | 5nF | Ci |
| 5nF |
2.3Pressure Equipment Directive
The MSP900SH transmitter does not enclose a pressurised volume and therefore is a component and outside the scope of the Pressure Equipment Directive 97/ 23/EC.
Accordingly, the Declaration of Conformity for the MSP900SH does not list the Pressure Equipment Directive.
2.4 | Specifications |
|
|
Materials of construction: |
|
| |
| Body Material | UPVC (stabilised) |
|
| Cable sealant | Epoxy adhesive |
|
| Locknut | Nylon |
|
| Bracket | 316 Stainless Steel |
|
| Cable | PVC sheathed two core shielded cable |
|
Electrical |
|
| |
| Supply voltage | Transmitter in Non Hazardous area: | |
|
| ||
|
| Transmitter in Hazardous area: | |
| Output |
| |
| Communications | HART Digital communications (Rev. 5) |
|
| Earthing | None required |
|
| Cable size | Overall diameter 4mm, two cores each 0.22 mm2 | |
| Cable length | 3m, 20m or 50m. |
|
| Cable resistance | 0.1 Ohm per metre length. |
|
Operating |
|
| |
| Range | 0.3 to 12m |
|
| Temperature |
|
|
| Ambient |
| |
| Wetside |
| |
| Pressure | 0 bar to 3.0 bar |
|
|
| (restricted to 0 - 1.0 bar for use in Canada) |
|
Page 4