Exergen DT 1001-LT, DT 1001-RS, DT 1001-LN Care and Maintenance Handling, Calibration, Cleaning

Models: DT 1001-LN DT 1001-RS DT 1001 DT 1001-LT

1 34
Download 34 pages 25.43 Kb
Page 9
Image 9

Care and Maintenance

Handling

Your DermaTemp is designed and built to industrial durability stan- dards in order to provide long and trouble-free service. However, it is also a high precision optical instrument, and should be ac- corded the same degree of care in handling as you would provide other precision optical instruments, such as cameras or otoscopes.

Calibration

Factory calibration data is installed via a computer through an op- tical link with the microprocessor. The instrument automatically self-calibrates each time it is turned on using this data, and will never require recalibration. If readings are not correct, the instru- ment should be returned for repair.

Cleaning

The DermaTemp can be gas or plasma sterilized, or wiped down with any hospital approved disinfectant, even bleach. With nor- mal use, the only maintenance required is to keep the lens on the end of the probe clean. It is made of special mirror-like, infrared- transmitting material called Germanium.

Dirt, greasy films or moisture on the lens will interfere with the passage of infrared heat and affect the accuracy of the instru- ment. If necessary, clean the lens with an alcohol prep or a cotton swab dipped in alcohol. Periodic cleaning is a good practice.

Self Diagnostics

Continuous Single Beeping

The high performance DermaTemp continuously monitors its abil- ity to produce accurate temperature readings. If either the target’s temperature or the unit’s ambient temperature exceeds the opera- tional limits, the beeper will sound once per second and the LED display will default to a display message.

When the target temperature is outside of the instrument’s oper- ating range, the unit will display either HI or LO and will beep con- tinuously at one beep per second. When the instrument’s own temperature is outside operating limits for ambient temperature, the display will show either HI A or LO A, and will beep continu- ously at one beep per second.