SPECIFICATIONS
Dimensions: | 140 mm x 173 mm x 135 mm (without chamber fitted) | |||
Weight: | 2.8 kg (no chamber fitted), 3.1 kg (chamber fitted and filled with water) | |||
Supply Frequency: | 50/60 Hz | Supply Voltage: | 115 V~ | Supply Current: 2.0 A max at 115 V~ |
Heater Plate: | 150 W | Heater Wire: | 22 V~, 2.73 A, 60 W, 50/60 Hz | |
Heater Plate |
|
| ||
Temperature Control Settings |
|
|
| |
Invasive Mode: | Chamber outlet: 35.5 - 37 °C, Airway 35 - 40 °C | |||
Noninvasive Mode: | Chamber outlet: 31 °C, Airway 28 - 34 °C | |||
Display: | Three digit 14 mm 7 segment LED |
| ||
Range: | 10 - 70 °C, Accuracy: ± 0.3 °C (in 25 - 45 °C temperature range) |
Alarm Parameters
High Humidity Alarm: An immediate, audible alarm at a displayed temperature of 41 °C or if the airway temperature exceeds 43 °C
Low Humidity Alarm: An audible alarm between 10 minutes @ 29.5 °C, and 60 minutes @ 34.5 °C (Invasive Mode only)
Sound Pressure Level: Alarms exceed 50 dBA @ 1 m
Performance
Recommended ambient temperature range: 18 - 26 °C
CAUTION: If operating in ambient temperatures outside the recommended range, consult your local Fisher & Paykel Healthcare representative.
Recommended Flow Range: Invasive Mode: up to 60 L/min, Noninvasive Mode: up to 120 L/min
|
| Refer to breathing circuit specifications for minimum flow | ||||
Humidity Performance: | Invasive Mode: > 33 mg/L, Noninvasive Mode: >10 mg/L | |||||
Maximum Operating Pressure: Refer to chamber and breathing circuit specifications | ||||||
| Less than 30 minutes |
|
| |||
Symbol Definitions |
|
|
|
|
| |
|
|
|
|
|
| |
|
|
|
|
|
|
|
Type BF | ATTENTION | Alternating | Drip | CAUTION | Respiratory Humidifier Classified by | |
| Consult | Current | Proof | Hot surfaces | Underwriters Laboratories Inc. with respect to | |
| accompanying |
|
|
| may exceed | electric shock, fire and mechanical hazards |
| documents |
|
|
| 85 °C | only in accordance with UL |
|
|
|
|
|
| CAN/CSA C22.2 No. 601.1 84DJ |
WARNING
Ensure that invasive mode is set for patients that have bypassed airways.
The use of breathing circuits, chambers or other accessories which are not approved by Fisher & Paykel Healthcare may impair performance or compromise safety.
Ensure that both temperature probe sensors are correctly and securely fitted. Failure to do so may result in temperatures in excess of 41 °C being delivered to the patient.
Ensure maintenance of grounding integrity by connection to a "hospital grade" receptacle. Always disconnect supply before servicing.
When mounting a humidifier adjacent to a patient ensure that the humidifier is always positioned lower than the patient.
The operation of high frequency surgical apparatus, shortwave or microwave equipment in the vicinity of the humidifier may adversely affect its function. If this occurs, the humidifier should be removed from the vicinity of such devices.
Do not touch the glass tip of the chamber temperature probe during use. Keep black connectors dry at all times.
Visually inspect accessories for damage before use.
CLEANING MR850 Heaterbase Using a damp cloth, clean the humidifier with either of the following: Isopropyl Alcohol, normal dishwashing detergent. Temperature
Probe Always clean before use. Probes can be cleaned with one of the following solutions: SporicidinTM, CidexTM OPA, SporoxTM OR sterilize the probe using Ethylene Oxide sterilization at 55 °C (131 °F), 80 kPa, (allow at least 15 hours for residual ETO to disperse before use). Wipe the airway temperature probes clear of any cleaning residues before use. Store probes in clean conditions. DO NOT autoclave probes.
CAUTION: DO NOT immerse the heaterbase or temperature probe electrical connections in any liquid.
NOTE: Follow the cleaning agent manufacturers’ instructions carefully. It is the user’s responsibility to qualify any deviations from these procedures, both for disinfecting efficacy and physical effect on the probe and the heaterbase. See the MR850 technical manual for more detailed instructions on cleaning.
MR850 RESPIRATORY HUMIDIFIER
Patents Pending in Principal Countries
of the World
The MR850 respiratory humidifier is used to warm and humidify gases delivered to patients requiring mechanical ventilation, positive pressure breathing assistance, or other medical gases.
INSTRUCTION SHEET
DO NOT DISCARD
REF 185042343
Rev G 2004-02
INTERNATIONAL P O Box 14 348, Panmure, Auckland 1134, New Zealand Tel:+64 9 574 0100 Fax:+64 9 574 0158 Email: info@fphcare.com Web Site: www.fphcare.com