NON-HEATED BREATHING CIRCUITS SET UP
1.Slide humidification chamber (a) onto humidifier base and connect breathing circuit (b) (refer to humidification chamber and breathing circuit operating instructions for further details).
2.Connect the temperature probe plug (c) (REF 900MR86X) to the blue socket on the humidifier base until an audible click is heard.
3.Push the chamber probe (d) and airway probe (e) into the breathing circuit. Make sure the chamber probe is correctly located in its
4. The humidification system is now set up and ready for use. After power on, the humidifier will default to invasive mode.
ATTENTION: Refer to operating instructions which accompany each accessory.
RECOMMENDED SETUP FOR ADULT PATIENTS900MR869 Temperature Probe, Mounting bracket to suit ventilator
For all other applications refer to Product Guide or contact your local Fisher & Paykel Healthcare representative.
SPECIFICATIONS (additional)
Temperature Control Settings (non heated circuits)Invasive Mode: Airway 37 ºC
Noninvasive Mode: Airway 31 ºC
Alarm Parameters (non heated circuits)High Temperature Alarm: Immediate, audible alarm at a displayed temperature > 41 ºC
Low Temperature Alarm: Audible alarm at a displayed temperature
<29 ºC (Invasive), < 26 ºC (Noninvasive)
Performance (non heated circuits)Recommended Flow Range: Invasive Mode: 6 to 60 L/min
(a) | (d) | (e) |
(b) |
To set the humidifier to
If a heated breathing circuit is used while in
The airway probe out test does not work with
When there is no gas flow the humidifier will give a low humidity alarm.
If the chamber runs dry either a water out alarm or a low humidity alarm will occur.