Stik Thermometer
SCPI Commands
Battery Life
Battery life is approximately 300 hours (12.5 days) of continuous operation with the backlight off. A
Calibration
For calibration information, refer to the document on the documentation CD.
SCPI Commands
For a list of the SCPI commands, refer to the document on the documentation CD.
RS-232 Interface
An
XWWarning
To prevent possible electrical shock, fire, or personal injury, the
Specifications
(Ambient: 23 °C ±5 °C) |
|
Measurement Range |
|
1551A Ex | −50 °C to 160 °C (−58 °F to 320 °F) |
1552A Ex | −80 °C to 300 °C (−112 °F to 572 °F) |
Accuracy (1 year) | ±0.05 °C (0.09 °F) |
Resolution | Selectable (0.1, 0.01, 0.001) factory default is 0.01 |
Sample Rate | User selectable 0.5/sec, 1/sec or 2/sec factory default is |
| 1/sec |
Operating Temperature Range of Readout: | −10 °C to 50 °C (14 °F to 122 °F) |
Probe Response Time | Approximately 20 seconds |
EMC Compliance | EN61326:2006 Annex C |
| CISPR 11, Edition |
| Class “B” |
Humidity Range | 0 to 95 % RH |
Storage Temperature Range | −20 °C to 60 °C (−4 °F to 140 °F) |
Enclosure Rating | IP50 |
Readout Temperature Coefficient | Add ±10 ppm/°C of full scale temperature from −10 °C to |
| 18 °C and 28 °C to 50 °C |
Probe Temperature Coefficient | 0.00385 Ω/Ω/°C nominal |
Nominal Probe Resistance at 0 °C | 100 Ω |
Probe Hysteresis | ±0.01 °C |
Power | 3 AAA alkaline batteries |
| (Must use only approved batteries. See Table 2) |
13