Fluke 2470 specifications Accuracy, Humidity Range

Models: 2470

1 78
Download 78 pages 56.11 Kb
Page 12
Image 12

RUSKA 2470

Users Manual

Temperature Range

Operating

Storage

Humidity Range

18 ºC to 28 ºC

-40 ºC to 70 ºC when thermometer and low range piston & Cylinder are stored separately from each other.

Operating

Storage

Pressure Medium

Pressure

Maximum Working Pressure

20% to 75% noncondensing

0% to 90% noncondensing

Clean dry gas, Nitrogen or equivalent, regulated to a pressure Compatible with each particular Piston/cylinder assembly. Dew Point of less than or equal to -60 ºF

2470 piston/Cylinder installed in 2470 Column adapter

With High Range Piston/Cylinder

3000 psig

2465 piston/Cylinder installed in 2465 Column adapter

With Mid Range Piston/Cylinder

100 psig (Do Not Exceed 6.31 Kg Mass Load)

With Low Range Piston/Cylinder

25 psig (Do Not Exceed 6.31 Kg Mass Load)

2468 piston/Cylinder installed in 2465 Column adapter

With Mid Range Piston/Cylinder

200 psig (Do Not Exceed 12.31 Kg Mass Load)

With Low Range Piston/Cylinder

50 psig (Do Not Exceed 12.31 Kg Mass Load)

Note

The pressurized housing with the 2470 adapter has been tested to 4000 psig. The pressurized housing with the 2465 adapter has been tested to 1000 psig.

Accuracy

The gauge is capable of measuring pressures to the accuracy indicated below. See the calibration report for the actual accuracy of this gauge. The claim for accuracy is valid only when the gauge is operated according to the instructions provided with the equipment. In addition, the claim is valid when the value of gravity acting upon the weights is known to +/-0.001 cm sec2.

Accuracy is defined as the departure of the measured pressure from the true pressure. The value is based on a simple error analysis of the calibration experiment and represents the sum of the systematic errors and two standard deviations of the random variability of the measurement process.

Gauge Pressure (0.0035% Class)

High Range Piston

0.0035 percent of reading

(0.013 sq in area)

or 0.002 psi, whichever is greater

Mid Range Piston

0.0035 percent of reading

(0.13 sq in area)

or 0.0002 psi, whichever is greater

Low Range Piston

0.0035 percent of reading

(0.52 sq in area)

or 0.00005 psi, whichever is greater

1-4

Page 12
Image 12
Fluke 2470 specifications Accuracy, Humidity Range