Specifications continued
Trend recording
Method |
|
| AutoTrend automatically records min, max and average values over time for all readings being | ||
|
|
|
|
| displayed for the 3 phases and neutral simultaneously |
Volts/amps/hertz, harmonics, power |
|
|
| and energy, flicker and unbalance mode | |
Sampling |
|
| 5 readings/sec continuous sampling per channel | ||
Memory |
|
| 1800 min, max and avg points for each reading | ||
Recording time |
|
| From 30 min with 1 second display resolution up to 450 days with 6 hour display resolution | ||
Zoom |
|
| Up to 6 x horizontal zoom | ||
Dips and swells mode |
|
|
|
|
|
Sampling |
|
| 100/1202 readings/sec continuous sampling per channel | ||
Memory |
|
| 3600 min, max and avg points for each reading | ||
Recording time |
|
| From 90 sec with 25 msec display resolution up to 450 days with 3 hour display resolution | ||
Zoom |
|
| Up to 12 x horizontal zoom | ||
Inrush currents and flicker PF5 mode |
|
|
| ||
Sampling |
|
| 100/1202 readings/sec continuous sampling per channel | ||
Memory |
|
| 3600 min, max and avg points for each reading | ||
Recording time |
|
| From 7.5 sec with 25 msec display resolution up to 30 min with 500 msec display resolution for | ||
|
|
|
|
| Inrush measurements and up to 2.5 hour with 2.5 sec display resolution for PF5 recordings |
Zoom |
|
| Up to 12 x horizontal zoom | ||
Monitor mode |
|
|
| ||
Sampling |
|
| Combination of 5 readings/sec and 100/1202 readings/sec continuous sampling per channel | ||
|
|
|
|
| depending on the parameter measured |
Recording time |
|
| Up to 1 week with 10 min resoluton | ||
Memory |
|
| 1008 min, max and avg points for each reading | ||
Limits |
|
| According EN50160 or customer definable | ||
Measurement method |
| ||||
Vrms, Arms |
|
| 10/122 cycle contiguous non overlapping intervals using 500/4162 samples per cycle in | ||
|
|
|
|
| accordance with IEC |
Vpeak, Apeak |
|
| Absolute highest sample value within 10/122 cycle interval with 40 µs sample resolution | ||
V Crest Factor |
|
| Measures ratio between the Vpeak and Vrms | ||
A Crest Factor |
|
| Measures ratio between the Apeak and Arms | ||
Hz |
|
| Measured every 10 sec in accordance with | ||
Vrms1⁄2 , Arms1⁄2 |
|
| Value is measured over 1 cycle, commencing at a fundamental zero crossing, and refreshed each | ||
|
|
|
|
| |
Harmonics |
|
| Calculated from | ||
|
|
|
|
| according to IEC |
Watt |
|
| Selectable Total or Fundamental real power display | ||
|
|
|
|
| Calculates average value of instantaneous power over 10/12 cycle period for each phase |
|
|
|
|
| Total Active Power PT =P1 + P2 + P3 |
VA |
|
| Selectable Total or Fundamental apparent power display | ||
|
|
|
|
| Calculates apparent power using Vrms x Arms value over 10/12 cycle period |
|
|
|
|
| Total Apparent Power is root mean square of real and apparent power |
VAR |
|
| Selectable Total of Fundamental reactive power display | ||
|
|
|
|
| Calculates reactive power as root of VA squared minus Watt squared over 10/12 cycle period |
|
|
|
|
| Capacitive and inductive load is indicated with capacitor and inductor icons |
Power Factor |
|
| Calculated Watt / VA | ||
Cos ϕ / DPF |
|
| Cos of angle between fundamental voltage and current | ||
|
|
|
| ||
Unbalance |
|
| The supply voltage unbalance is evaluated using the method of symmetrical components | ||
|
|
|
|
| according to |
Flicker |
|
| According to IEC | ||
|
|
|
|
| Includes 230 V 50 Hz lamp and 120 V 60 Hz lamp models |
Transient capture |
|
| Captures waveform triggered on signal envelope | ||
|
|
|
|
| Additionally triggers on dips, swells, interruptions and Amps level as specified by |
Inrush current |
|
| The inrush current begins when the Arms half cycle rises above the inrush threshold, and ends | ||
|
|
|
|
| when the Arms half cycle rms is equal to or below the inrush threshold minus a |
|
|
|
|
| hysteresis value. The measurement is the square root of the mean of the squared Arms half |
|
|
|
|
| cycle values measured during the inrush duration. Each |
|
|
|
|
| |
|
|
|
|
| Cursors allow measurement of peak Arms half cycle. |
6 Fluke Corporation Fluke 430 Series