Specifications continued

General

Case

Design

Rugged, shock proof with integrated protective holster

Drip and dust proof

IP51 according to IEC60529 when used in tilt stand position

Shock and Vibration

Shock 30 g, Vibration: 3 g Sinusoid, Random 0.03 g2/Hz according to MIL-PRF-28800F Class 2

Display

 

 

 

Type

Bright Full-Color LCD with CCFL backlight, 80cd/m2

Size

115.2 x 86.4 mm (4.5 x 3.4 in)

Resolution

320 x 240 pixels

Contrast and brightness

User adjustable, temperature compensated

Memory

 

 

Screens

50 screen memories on Fluke 434

 

 

 

25 screen memories on Fluke 433

Data

10 data memories for storing data including recordings on Fluke 434

 

 

 

5 data memories for storing data including recordings on Fluke 433

Limit templates

2 preprogrammed, 2 administrator (programmable via FlukeView, 2 user locations)

Real-time clock

Time and date stamp for AutoTrend, Transient display and SystemMonitor

Mechanical

 

Size

256 x 169 x 64 mm (10 x 6.6 x 2.5 in.)

Weight

2 kg (4.4 lb.)

Power

 

Line power

Switchable 115 V, 230 V adapter with country specific plug

Battery power

Rechargeable NiMH (installed)

Battery operating time

> 7 hours

Battery charging time

4 hours

Power saving

Adjustable time for dimmed backlight with on screen power indicator

Standards

 

Measurement methods used

IEC61000-4-30 class A

Power Quality

EN50160

Flicker

IEC 61000-4-15

Harmonics

IEC 61000-4-7

Safety

 

Compliance

IEC/EN61010-1 (2nd edition) pollution degree 2; CAN/CSA C22.2 No 101.1; ANSI/ISA S82.01

Max voltage on banana input

CAT III 1000 V / CAT IV 600 V

Max voltage on current BNC input

42 Vpeak

Environmental

 

Operating temperature

0 °C to +50 °C (32 °F to 122 °F)

Storage temperature

-20 °C to +60 °C (-4 °F to 140 °F)

Humidity

10 to 30 °C: 95 % RH non condensing (50 to 86 °F: 95 % RH non condensing)

 

 

 

30 to 40 °C: 75 % RH non condensing (86 to 104 °F: 75 % RH non condensing)

 

 

 

40 to 50 °C: 45 % RH non condensing (104 to 122 °F: 45 % RH non condensing)

Maximum operating altitude

3000 m (9842 ft) Derate to CAT II 1000 V / CAT III 600 V / CAT IV 300 V above 2000 m (6561 ft)

Maximum storage altitude

12 km (39,372 ft)

Warranty

3 years on mainframe, 1 year on included accessories

Accessories

 

Included

Hard carrying case with clamp holders BC430

 

 

 

4 current clamps, i400s

 

 

 

5 Test leads, 4 black, 1 green

 

 

 

5 Alligator clips, 4 black, 1 green

 

 

 

Battery Charger Eliminator, BC430

 

 

 

FlukeView Software, SW43W (Fluke 434 only)

 

 

 

Optical Cable for USB , OC4USB (Fluke 434 only)

 

 

 

Color localization set, WC100

 

 

 

Getting Started (printed)

 

 

 

User Manual (CD-ROM)

1depending clamp scaling

250 Hz/60 Hz nominal frequency according to IEC 61000-4-30

3Add clamp accuracy

4The advanced functions: interharmonics, energy, transients and inrush are optional for the Fluke 433 and standard available on the Fluke 434

Fluke 430 Series Three-Phase Power Quality Analyzers Fluke Corporation 7