50S & 50D Thermometer

Calibration Information

5.Allow the system to stabilize, then adjust T1 or T2 offset adjustments for a temperature display equal to the reading of the lag bath reference thermometer.

6.Calibration of the thermometers is complete. Disconnect all test equipment from the instrument.

50S or 50D

 

Thermometer

 

+

_

 

 

Thermocouple wire

Mercury

 

 

 

 

thermometer

 

 

(0.02°C resolution)

 

 

Lag bath

 

 

(room temperature

 

 

water)

 

 

Immersed to

 

 

approximately

 

 

same depth

 

 

zf06f.eps

Figure 7. Room Temperature Lag Bath