50S & 50D Thermometer
Calibration Information
5.Allow the system to stabilize, then adjust T1 or T2 offset adjustments for a temperature display equal to the reading of the lag bath reference thermometer.
6.Calibration of the thermometers is complete. Disconnect all test equipment from the instrument.
50S or 50D |
| |
Thermometer |
| |
+ | _ |
|
| Thermocouple wire | Mercury |
|
| |
|
| thermometer |
|
| (0.02°C resolution) |
|
| Lag bath |
|
| (room temperature |
|
| water) |
|
| Immersed to |
|
| approximately |
|
| same depth |
|
| zf06f.eps |
Figure 7. Room Temperature Lag Bath