50S & 50D Thermometer
Calibration Information
50S or 50D | Thermocouple wire |
Thermometer | + |
| _ |
| Ice bath |
| 5520A |
| Copper wire |
| zf05f.eps |
Figure 4. Ice Bath Equipment Connections
Ice Bath Test
Use the following procedure to test the thermometer. This test should be conducted on both thermocouple inputs of the 50D.
1.Connect the equipment as shown in Figure 4.
2.Insert the Mercury Thermometer to the same depth as the thermocouple wires and verify that the ice bath temperature is 0.0 oC ± 0.1 oC.
3.Short (or apply 0 mV) to the copper wires at the 5520A Calibrator.
4.Verify the thermometer reads 0.0 oC ± 0.4 oC.
5.If this test fails, reposition the thermocouple wires and Mercury Thermometer in the ice bath and repeat steps
6.Remove the short applied in step 3.
7.Using Table 3 (for the corresponding thermocouple type), obtain an output from the 5520A equal to the first value of the “Input Voltage (mV DC)” column, and verify that the thermometer reads within the limits specified.
8.Repeat these steps for the remaining voltages in the “Input Voltage (mV DC)” column.
9.Disconnect the thermocouple from the input terminals. This completes the ice bath test.
8