51, 52, 53, 54 Series II
Service Manual
Specifications
General Specifications
Weight
Dimensions (without holster)
Battery
Certification
280 g (10 oz)
2.8cm ´ 7.8 cm ´ 16.2 cm (1.1 in ´ 3 in ´ 6.4 in)
3 AA batteries
P, )(TÜV pending)
Safety | CSA C22.2 No. 1010.1 1992 |
| EN 61010 Amendments 1, 2 |
CAT I | OVERVOLTAGE (Installation) CATEGORY I, |
| Pollution Degree 2 per |
*Refers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient over voltages to an appropriate low level.
Environmental
Operating Temperature
Storage Temperature
Humidity
Non Condensing (< +10 oC) (< +50 oF)
95% RH (+10 oC to +30 oC) (+50 oF to +86 oF) 75% RH (+30 oC to +40 oC) (+86 oF to +104 oF) 45% RH (+40 oC to +50 oC) (+104 oF to +122 oF) (Without Condensation)
Electrical
Measurement Range
Display
Resolution
Measurement Accuracy
Temperature
Coefficient
Electromagnetic Compatibility
Maximum
Differential
Common Mode
Voltage
Temperature
Scale
Applicable
Standards
R- and
0.1 oC / oF / K < 1000o 1.0 oC / oF / K ³ 1000o
[below
R- and
0.01 % of reading + 0.03 oC per oC
(0.05 oF per oF) outside the specified +18 oC to +28 oC (+64 oF to +82 oF) range [below
Susceptibility: ± 2 oC (± 3.6 oF) for 80 MHz to 200 MHz in 1.5 v/m field, for 200 MHz to 1000 MHz in 3 V/m field.
Emissions: Commercial limits per
1 V (Maximum voltage difference between T1 and T2)
Accuracy is specified for ambient temperatures between +18 oC (+64 oF) and +28 oC (+82 oF) for a period of 1 year. The above specifications do not include thermocouple error.