Specifications
Specifications
Measurement Range
Operating Temperature (Ambient)
Storage Temperature
Display Resolution
Measurement Accuracy
Temperature Coefficient
Response Time
Spectral Response
Field of View/Target Size
Laser beam divergence
Repeatability
Emissivity
Humidity
Altitude
Sensor Element
Power Supply
Battery Life
Backlight
Hold
Dimensions
Weight
Safety/Regulatory Compliance
−40 °C to +500 °C ( −40 °F to +932 °F)
0 °C to +50 °C ( +32 °F to +122 °F)
−20 °C to +70 °C ( −4 °F to +158 °F), without battery
Below 200 °: 0.1°
Above 200 °: 1°
Below 0 °C: ±5 °C (32 °F: ±9 °F)
Above 0 °C: ±2 °C (32 °F: ±4 °F)
Above 100 °C: ±2 % of reading
[Ambient: 25 °C ±3 °C (77 °F ±5 °F)]
>400 °C (752 °F): ±0.06 % reading
≤400 °C (752 °F): ±0.24 % reading [Ambient: < 22 °C (72 °F); > 28 °C (82 °F)] 0.8 second (for 95 % response)
8 μm to 14 μm nominal
8:1; 25 mm (1 in) minimum spot size
<0.01 radian
±1 % of reading or ±1 °, whichever is greater
Fixed at 0.95
10% to 90 % RH
2AA alkaline batteries, not installed
>15 hours with laser and backlight activated (> 4000 individual measurements under typical conditions)
Temperature value is held on the screen for 20 seconds
38.1mm H × 63.5 mm W × 185.4 mm L
(1.5 in H × 2.5 in W × 7.3 in L)
283.5 g (10 oz)
Certification
Approval
US 21 CFR Subchapter J Part 1040.10
IEC
EN
3