Model 65
Calibration Information
8.Press D to enter the
9.Press G. The default
10.Aim the UUT at the blackbody target; then use F or D to adjust the UUT reading down or up, respectively, to read 350 °C.
11.When you have set the UUT reading to 350 °C, press G. The UUT display flashes briefly, then returns to the calibration selection mode.
Note
If the high temperature you enter during calibration varies greatly from the measured temperature, the UUT display will flash Err after you press G. In this case, press G again to exit the error mode. Verify that the blackbody is working properly; then try the calibration again.
12.Press D to enter the
13.Test the UUT’s operation by running the performance tests given earlier under “Performance Tests”. If the UUT fails the performance tests, recalibrate the UUT and try the tests again. If the UUT continues to fail, return it to a Fluke Service Center for repair or replacement.
Replacement Parts
Table 3 shows the replacement parts available from Fluke for the Model 65 Infrared Thermometer.
Table 3. Replacement Parts
Description
Top case
Bottom case
Battery door
LCD lens
End cap
Keypad
Softcase
Holster
AAbattery, 1.5 V,
Model 65 Infrared Thermometer Instruction Sheet
Americas (English, Spanish, French, Portuguese)
European I (English, French, German, Italian, Finnish, Dutch, )
European II (English, Norwegian, Spanish, Portuguese, Swedish, Danish)
Asian (English, Korean, Japanese, Simplified Chinese, Traditional Chinese, Thai)
Model 65 Infrared Thermometer Quick Reference Card (English only)
Fluke Part Number
803566
803509
803517
803558
803525
803541
670676
670612
650181
376756
801826
801842
801859
801834
802261
8