80K-15
Electronic Air Cleaner Probe
Symbols
TDouble insulated.
WRefer to explanation in this instruction sheet.
XLethal voltages may be present.
Complies with relevant Canadian Standards
4.Connect the alligator clip of the probe’s grounding lead to earth ground.
5.Touch the probe tip to the circuit being measured and read the measurement on the meter display. If necessary, apply a correction factor to the reading. (See “Note” in the “Specifications,” below.
Table 1. 80K-15 Circuit Loading and Input/Output
Association directives.
Earth ground
Using the Probe
Input Voltage
10 V |
Characteristics
Loading Current
10 nA
Output Voltage
10 mV |
1 |
87 III TRUE RMS MULTIMETER
|
| DC |
MIN MAX | RANGE | HOLD |
| REL | Hz |
41/2 DIGITS PEAK MIN MAX |
|
|
1 Second |
|
|
mV |
|
|
100 V |
1 kV |
10 kV |
15 kV |
100nA
1μA
10μA
15μA
100 mV |
1 V |
10 V |
15 V |
2 |
V | mA |
A |
V∝A
OFF
A mA∝A COM V
! |
| 400mA MAX | CAT II |
10A MAX | FUSED | 1000V MAX |
FUSED | ! |
|
3 |
4 |
5 |
Specifications
The
The accuracy of the probe does not include the accuracy of the meter. To get the accuracy of the system, the accuracy of the meter must be added to the accuracy of the probe.
Note
The input impedance of the Autoranging mode on Fluke handheld digital multimeters varies by range.
EAOO1F.EPD
AProbe tip
BProbe body
CPlug
DGND tab
EGrounding lead clip
FMultimeter or voltmeter
Figure 1. 80K-15 Electronic Air Cleaner Probe
The input impedance in almost all ranges of Fluke multimeters is ~10 MΩ. The exceptions are: the 3 V range on the Fluke Models 21, 23, 25, 27, 70, 73, 75, 7; and the 4 V range on the Models 10, 11, 12, 16, 79, 83, 85, 86, 87, 88. In the ranges on these multimeters, the impedance is 11.11 MΩ.
To improve the measurement accuracy of the probe when using these ranges on the models indicated, apply a correction factor of 0.99 i.e., multiply the display reading by 0.99.
The
The
Taking care to follow the safety practices under “WRead First: Safety Information,” use the probe as follows:
1.Turn on a compatible voltmeter or multimeter and select the voltage measurement function.
2.Select an appropriate voltage range (i.e., 1 volt reading per 1000 volt input. See Table 1).
3.Connect the probe’s output leads to the voltmeter or multimeter input terminals.
Voltage Range:
Maximum Current:
Input Resistance:
Division Ratio:
Accuracy DC:
Accuracy AC:
Safety:
Altitude:
1 kV to 15 kV dc or peak ac,
10 kV rms ac
20μA 1000 MΩ
1000 : 1 (1000 x attenuation)
±2 % in 10 o C to 45 o C
±5 % @ 60 Hz, in 20 o C to 30 o C
Complies with IEC
2000 meters