Fluke warranty Symbols, Using the Probe, Specifications, 80K-15Circuit Loading and Input/Output

Models: 80K-15

1 3
Download 3 pages 34.94 Kb
Page 2
Image 2
80K-15

80K-15

Electronic Air Cleaner Probe

Symbols

TDouble insulated.

WRefer to explanation in this instruction sheet.

XLethal voltages may be present.

Complies with relevant Canadian Standards

4.Connect the alligator clip of the probe’s grounding lead to earth ground.

5.Touch the probe tip to the circuit being measured and read the measurement on the meter display. If necessary, apply a correction factor to the reading. (See “Note” in the “Specifications,” below.

Table 1. 80K-15 Circuit Loading and Input/Output

Association directives.

Earth ground

Using the Probe

Input Voltage

10 V

Characteristics

Loading Current

10 nA

Output Voltage

10 mV

1

Symbols 87 III TRUE RMS MULTIMETER

 

 

DC

MIN MAX

RANGE

HOLD

 

REL

Hz

41/2 DIGITS PEAK MIN MAX

 

 

1 Second

 

 

mV

 

 

100 V

1 kV

10 kV

15 kV

100nA

1μA

10μA

15μA

100 mV

1 V

10 V

15 V

2

V

mA

A

VA Table 1. 80K-15 Circuit Loading and Input/Output

OFF

A mAA COM VUsing the ProbeInput Voltage

!

 

400mA MAX

CAT II

10A MAX

FUSED

1000V MAX

FUSED

!

 

3

4

5

Specifications

The 80K-15 Electronic Air Cleaner Probe achieves its rated accuracy when it is used with a voltmeter or multimeter (ac or dc) with an input impedance of 10 MΩ ± 1%.

The accuracy of the probe does not include the accuracy of the meter. To get the accuracy of the system, the accuracy of the meter must be added to the accuracy of the probe.

Note

The input impedance of the Autoranging mode on Fluke handheld digital multimeters varies by range.

EAOO1F.EPD

AProbe tip

BProbe body

CPlug

DGND tab

EGrounding lead clip

FMultimeter or voltmeter

Figure 1. 80K-15 Electronic Air Cleaner Probe

The input impedance in almost all ranges of Fluke multimeters is ~10 MΩ. The exceptions are: the 3 V range on the Fluke Models 21, 23, 25, 27, 70, 73, 75, 7; and the 4 V range on the Models 10, 11, 12, 16, 79, 83, 85, 86, 87, 88. In the ranges on these multimeters, the impedance is 11.11 MΩ.

To improve the measurement accuracy of the probe when using these ranges on the models indicated, apply a correction factor of 0.99 i.e., multiply the display reading by 0.99.

The 80K-15 (Figure 1) is designed to work with a meter that has an input impedance of 10 MΩ ( ± 1% ). A meter with an input impedance less than 10 MΩ can produce readings lower than the voltage present. This can pose a serious hazard when a dangerous voltage is present.

The 80K-15 represents a 1000 MΩ load to the circuit being measured, or 1 μA per 1 kV. Table 1 shows the circuit loading and input/output characteristics of the probe over its measurement range.

Taking care to follow the safety practices under “WRead First: Safety Information,” use the probe as follows:

1.Turn on a compatible voltmeter or multimeter and select the voltage measurement function.

2.Select an appropriate voltage range (i.e., 1 volt reading per 1000 volt input. See Table 1).

3.Connect the probe’s output leads to the voltmeter or multimeter input terminals.

Voltage Range:

Maximum Current:

Input Resistance:

Division Ratio:

Accuracy DC:

Accuracy AC:

Safety:

Altitude:

1 kV to 15 kV dc or peak ac,

10 kV rms ac

20μA 1000 MΩ

1000 : 1 (1000 x attenuation)

±2 % in 10 o C to 45 o C

±5 % @ 60 Hz, in 20 o C to 30 o C

Complies with IEC 1010-2-31:1993, Type B, 15 kV dc or peak ac, 10 kV rms ac, Overvoltage Category I (voltages derived from limited energy transformer).

2000 meters

Page 2
Image 2
Fluke Symbols, Using the Probe, Specifications, 80K-15Circuit Loading and Input/Output, Input Voltage, Output Voltage