Fluke 943-27, 943-227-15 manual Configuration Functions, Firmware Version, Deleted, Not Used

Models: 943-227-15 943-27

1 45
Download 45 pages 38.43 Kb
Page 23
Image 23

943-227-15 Digital Preamplifier

Operator Manual

For UDR systems, the address of the 942-200-80 Communications board in the UDR must be set 4040 and the baud rate must be set to 4800 baud. If the wiring of the Communications option board to the P2 connector is suspected, a loop test can be performed as described in "Loop Test Mode". The instruction manual for the Communications option board (P/N 942-200-80-1) contains information on how to set up the serial board in the UDR.

Firmware Version

This system requires a PROM for operation. For the part number and the latest revision of the PROM, see the Factory Test Data Sheet. This Manual has been prepared for use with the P/N 227LNK, Rev. 2, PROM.

3.4 Configuration Functions

The preamplifier configuration jumpers and required settings for this application are listed below:

Table 3-2. Preamplifier Serial Interface Board 977-210-10M Switch Settings

Switch

Normal

Switch Function

 

Op.

 

SW1

Open

Microprocessor Reset switch, momentary contact

 

 

 

SW2

N/A

Not supplied

 

 

 

SW3-1

ON

OFF Selects 300 baud • 1/ON Selects 4800 baud (normal setting)

SW3-2

OFF

OFF • (No function is implemented in the 977-210-10M)

SW3-3

OFF

OFF Selects Rate Mode (normal) • (Selects Maintenance Mode for

 

 

factory PGA and Auto Zero set-up only)

SW3-4

OFF

OFF • (No function is implemented in the 977-210-10M)

NOTE

SW3 is a 4-position DIP switch on the 977-210-10M serial interface board.

Table 3-3. Jumper Positions for Preamplifier Serial Interface Board 977-210-10M

Jumper

Normal

Function

Alternate

Function

 

Setting

 

Setting

 

JMP1

A to B

Enables Victoreen Loop

B to C

Enables RS232C

 

 

(normal)

 

 

JMP2

N/A

Deleted, Not Used

N/A

Deleted, Not Used

Table 3-4. Jumper Positions for Preamplifier Electrometer Board 977-200-15M

Jumper

Normal

Function

Cal.

Function

 

Setting

 

Position

 

JMP1

A to B

Normal Operation

B to C

PGA Offset Adjust *

* Programmable Gain Amplifier requires a special test PROM and is a factory adjustment

3-5

Page 23
Image 23
Fluke 943-27, 943-227-15 Configuration Functions, Firmware Version, To B Enables Victoreen Loop To C Enables RS232C Normal