Specifications and Applications

Image Capturing Unit (Camera Head and Controller)

CCD chip

Fujifilm Super CCD Area Type chip

 

Number of pixels

3.2M pixels

 

Pixel size

10.75 x 10.75µm

 

Cooling

Two-stage thermoelectric module with air circulation

Cooling temperature

-30°C (when room is below 28°C)

 

Focusing

Power focusing; remote and preset control

 

Exposure time

1/100 second to 30 hrs (2 hrs to 30 hrs to be set manually)

Dynamic range

Four orders of magnitude

 

Gradation

16 bits (65,536)

 

Image size

12.6MB Max.; 49.2KB Min.

 

Maximum sample size

25 x 25cm (wide angle lens); 14 x 21cm (Fujinon VRF43LMD lens)

Binning

1 x 2, 2 x 4, 4 x 8, and 8 x 16 pixels

 

Interface

USB1.1

 

Software

 

 

Image capture

Fujifilm Image Reader (Macand Windows®)

Image analysis

Fujifilm Image Gauge (Mac); Fujifilm MultiGauge (Windows®)

Dimensions and Weights

 

 

Camera head

180 (W) x 170 (H) x 250mm (D)

3.4kg

Dark Box IV

510 (W) x 730 (H) x 480mm (D)

49.0kg

Analyzing Unit

 

 

The binning mode of the LAS-3000 allows researchers to select from four binning settings to enhance both imaging sensitivity and image resolution.

Sample is placed on a tray

Chemiluminescence

S u p e r C C D

R e m o t e C o n t r o l l e d

Operating system

Windows® XP or MacOS 9 & X

Applicable Reagents and/or Samples

Chemiluminescence

CDP-Star®, ECL, ECL Plus, SuperSignal, ImmunoStar, CSPD®

Fluorescence

EtBr (W/UV light), SYBR® Green I & II, SYPRO® Orange,

 

GFP, DY-458XL (blue LED)

 

Pro-QDiamond, Cy3, RFP, DY-520XL (green LED), Alexa Fluor 633,

 

Cy5, DY-647 (red LED)

Chemifluorescence

*AttoPhos

Digitization

CBB-stained gel, Silver stained gel (Transilluminator)

 

NBT/BCIP-stained membrane. MW marker (Epi-illuminator)

Intelligent Dark Box IV

EPI-illuminator for fluorescence

blue LED (470nm), green LED (520nm), red LED (630nm)

EPI-illuminator for digitization

White-light source

 

Transilluminator - for digitization of

 

 

 

stained gels and autoradiographic films

White LED

 

UV transilluminator**

 

UV-light source (312nm)

 

Filter turret

 

Five positions

 

Printers

 

 

 

 

Pictrography 3500

 

 

 

 

Lens

 

 

Lens

 

High-sensitivity lens (FUJINON VRF43LMD)

 

Wide-view lens

 

F-number

0.85

 

F-number

2.0

Focal length

43mm

 

Focal length

24mm

Focus

Remote power focusing

Focus

Manual

Set a tray

Sample

Remote controlled

focusing

Exposure

Analysis

 

Chemiluminescence Mode

 

Fluorescence

 

 

Chemifluorescence

 

 

Filter

blue LED

UV trans-

green LED

illuminator

red LED

 

Fluorescence

 

 

Chemifluorescence Mode

 

Digitization

 

 

 

White light

White

trans-

epi-illuminator

illuminator

 

Digitize Mode

S c i e n c e I m a g i n g S y s t e m

Mount

Bayonet

Mount

Adapter to Nikon F mount

*No license is granted for use of AttoPhos™ to detect nucleic acid on a nylon membrane.

**No license is granted for pre-labeling gel electrophoresis method with the UV transilluminator.

The image analysis process captures images in three modes: chemiluminescence, fluorescence/chemifluorescence and digitization.

LAS-3OOO

Fuji Photo Film Co., Ltd. 26-30, Nishiazabu 2-Chome, Minato-ku, Tokyo 106-8620, Japan, Tel: +81-3-3406-2201, Fax: +81-3-3406-2158 • http://lifescience.fujifilm.com • E-mail: sginfo@tokyo.fujifilm.co.jp

(04,09)

Fujifilm Medical Systems U.S.A., Inc. 419 West Avenue, Stamford, CT 06902, U.S.A., Tel: +1-203-324-2000 ext. 6112 (1-800-431-1850 ext. 6112 in the U.S.), Fax: +1-203-351-4713 • http://www.fujimed.com • E-mail: SSG@fujimed.com

-204E

Fuji Photo Film (Europe) GmbH, Heesenstr. 31, 40549 Düsseldorf, Germany, Tel: +49-211-5089-174, Fax: +49-211-5089-139 • http://www.fujifilm.de • E-mail: mkaling@fujifilmeurope.de

No. BB

Specifications and system configuration subject to change for improvement without notice. All other product names mentioned herein are the trademarks of their respective owners.