GE 60-639-95R-OD, 60-639-43-EUR-OD Programming, Final Testing, Maintenance, Replacing Batteries

Models: 60-639-95R 60-639-43-EUR 60-639-43-EUR-OD 60-639-95R-OD

1 8
Download 8 pages 45.1 Kb
Page 6
Image 6
Programming

1.Determine which detection zone/lens segment needs a masking label.

2.Peel the desired mask label from its backing and apply to the inside of the lens segment to be blocked.

Programming

Refer to the panel installation instructions for information on programming the sensor into the

 

panel.

 

 

 

 

 

 

 

 

 

 

 

 

Final Testing

Final testing should be done to verify radio signal integrity and confirm control panel program-

 

ming and response. The actual transmitter range can be determined by performing a sensor test as

 

follows:

 

 

 

 

 

 

 

 

 

 

1. After the sensor has been mounted, remove it from its mounting plate and activate the tamper

 

switch to start the walk test mode.

 

2. Replace the sensor in its mounting plate.

 

3. Place the control panel in test mode. Move across the detection pattern until the sensor’s LED

 

turns on. STOP your motion.

 

4. Listen for the appropriate system response. If the system does not respond, proceed to the

 

“Troubleshooting” section.

 

 

 

Maintenance

At least once a year, the range and coverage should be verified for proper operation. The end user

 

should be instructed to put the sensor in walk test mode and walk through the far end of the cov-

 

erage pattern to verify proper detection.

 

Replacing Batteries

 

When battery replacement is necessary, observe proper polarity (as shown in the battery compart-

 

ment) when installing the new battery. The sensor may be damaged if installed incorrectly. As

 

you look at the battery compartment, on the left side the positive battery end is down and on the

 

right side the positive end is up. When the battery is replaced, wait at least 3 minutes after install-

 

ing the battery before activating the walk test mode. See Figure 8 for battery locations.

 

!

Replace only with two AA 1.5V Alkaline batteries. Dispose of used batteries

 

according to manufacturer’s instructions and local government authorities.

 

 

 

 

 

 

 

 

 

 

 

Caution

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PIR

 

 

2

AA

 

 

 

Cover

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ba

tt

1

.5 V

 

 

 

 

 

 

e

 

 

ol

 

 

 

 

 

ries

 

t

 

 

 

 

 

 

 

 

 

 

PWB

 

 

 

 

 

 

 

 

 

 

Cover

 

 

 

 

 

 

 

Lens

 

 

 

 

 

 

 

 

 

Tabs

Final Testing Sensor

Body

Tamper

Switch

Mounting

Plate

Figure 8. PIR Components

Troubleshooting Follow these guidelines if the system does not respond correctly when the sensor is activated.

Check programming and re-program sensor into panel if necessary.

Move the sensor to another location and test for correct response.

6

Indoor/Outdoor PIR Motion Sensors Installation Instructions

Page 6
Image 6
GE 60-639-95R-OD, 60-639-43-EUR-OD manual Programming, Final Testing, Maintenance, Replacing Batteries