TROUBLESHOOTING GUIDE

SYMPTOM

POSSIBLE CAUSES

REMEDIES

 

 

 

Low permeate flow rate

Dirty or fouled membrane

Flush and clean the membrane

(continued)

elements

elements.

 

 

 

 

Operating on cold water less

Install a hot/cold feed water

 

than 55ºF (13ºC)

tempering valve if more perme-

 

 

ate flow is needed. Operate

 

 

with a feed water temperature of

 

 

72º - 77ºF (22º - 25ºC).

 

 

 

 

Membrane elements installed

Install membrane elements in

 

backwards or damaged

the direction of fluid flow.

 

concentrate seal

Flush and clean the machine

 

 

immediately. Membrane ele-

 

 

ments with damaged concentrate

 

 

seals should be cleaned and may

 

 

be returned for repair.

 

 

 

 

Flow meter inaccurate

Check the flow rate manually

 

 

with a stopwatch and calibrated

 

 

container.

 

 

 

Low concentrate flow rate,

Concentrate valve plugged

Remove the concentrate valve

 

stem and/or disassemble the

normal or higher than

 

 

plumbing. Clean the valve.

normal pressure

 

 

 

 

 

 

 

Concentrate outlet line restricted

Examine the concentrate line for

 

 

obstructions or kinks, repair or

 

 

replace the tubing.

 

 

 

 

Flow meter inaccurate

Check the flow rate manually

 

 

with a calibrated container.

 

 

 

Pressure does not drop when

Dirty concentrate valve

Disassemble and clean the

concentrate valve opened

 

plumbing to the valve.

 

 

 

 

 

27