RO Power Switch:
The RO Power Switch is mounted on the back of the remote monitor module. The switch is used to turn the
Inlet Water Sensor:
Monitors the quality of the inlet water. Inlet quality is compared to the outlet (product) water quality and is displayed as "percent rejection" on the front panel.
Product Water Sensor:
The Product Water Sensor is used in conjunction with the inlet water sensor by the
Total Dissolved Solids (TDS) Sensor:
Samples the product water quality via the sensor probe, and the final quality of the product water is indicated on a digital display on the front panel as TDS in mg/L and ppm. The monitor is temperature compensated to adjust the display value to changing water temperature.
Pressure Adjust Regulator:
The Pressure Adjust Regulator inside the RO is used to set the maximum product pressure needed to supply a dialysis machine. Adjusting this regulator will also result in small adjustments in the pump pressure.
Pre and Post filter Pressure Gauges:
The
Product Pressure Gauge:
The Product Pressure Gauge reads the product pressure from 0 to 100 PSI and is located on the RO module right side panel. Product pressure will read zero (or close to) when the product line is not connected to a dialysis unit and the product hose is connected to the product fitting and the valve is open. The Fresenius dialysis machine requires a minimum inlet pressure of 20 PSI.
Pump Pressure Gauge:
The Pump Pressure Gauge reads the pump pressure from 0 to 300 PSI and is located on the RO module right side panel. It is used for monitoring during operation and when adjusting the pump pressure, and is located on the front of the RO.
12552e – 27 Feb 04 | 1 - 9 | General Information |