Millenium™ RO System by GE Osmonics
1.10Specifications:
1.10.1Input Water Requirements
| Minimum | Maximum |
Input flow Rate | 2.0 GPM | N/A |
Temperature (water) | 1.7°C (35°F) | 32°C (90°F) |
NOTE:
Operating at lower temperatures may make it impossible for the Millenium™ to produce adequate permeate for the dialysis machine.
pH (operating) | 4.0 |
| 11.0 |
TF Membranes |
|
|
|
pH w/chloramines present (operating) | 4.0 | 8.5 | |
TF Membranes |
|
|
|
pH (cleaning) |
| 2.0 | 11.0 |
TF Membranes |
|
|
|
Pressure, dynamic |
| 30 PSI | 100 PSI |
1.10.2Flow Rates
Product Flow based on the following conditions:
| New Membranes |
| ||
| Operating Pressure: | 190 PSI | ||
| Product Pressure: | 1 PSI | ||
| Feed Water Temperature: | 77°F (25°C) | ||
GPM | LPM | GPD | LPD |
|
.34 | 1.31 | 500 | 1892.5 |
|
.52 | 1.97 | 750 | 2839.1 |
|
Membrane flux may vary ±15%
Waste Flow: .84 GPM (3.2 LPM) (±10%)
1.10.3Electrical Requirements
Motor | 1/3 HP |
Voltage | 115 VAC / 230 VAC |
Hertz | 60 / 50 |
Amps | 8 / 5 |
Phase | single |
14693e – 31 Mar 03 | 1 - 12 | General Information |