4.3.1Pre-Cleaning Preparation

There are two types of cleaners available, a low pH cleaner (Calcium/Iron removal) and a high pH cleaner (Silt/Organics removal). GE Osmonics provides these cleaners in liquid form. The low pH cleaner attacks mineral scale deposits such as calcium, magnesium and iron build-up, and should be used first. The high pH cleaner will remove accumulated silt and organic matter from the RO membrane.

CAUTION:

The different cleaners must not be mixed together. Introduce the low pH cleaner first, thoroughly rinse the system, and then use the high pH cleaner.

WARNING:

The solution is a mild irritant to eyes and skin. It is recommended that protective gloves and eye protection be worn when handling the cleaner. Neither respiratory protection nor special ventilation are required.

Materials Required:

Liqui-Clean 403™ – (Low pH liquid) - GE Osmonics Part No. 1228653 (1 Gal) Liqui-Clean 211™ – (High pH liquid) - GE Osmonics Part No. 1228791 (1 Gal)

high and low pH cleaner

2 Gallons pretreated water

clean wide mouth container with volume of 2 ½ gallon or more

pH test strips

product rinse line (clear hose with male QD fitting on one end, open on the other)

“DO NOT USE” warning labels (not provided by Osmonics)

cleaning / disinfect log sheets

most current log sheets for water quality comparison

4.3.1.1Pre-Cleaning Procedure

Read and understand instructions on cleaner container before beginning procedure.

1.Connect the water inlet line to the water supply, and the waste and product lines to drain. Turn the RO unit and water supply on.

2.Establish a pre-cleaning % rejection, TDS and pH record on the cleaning disinfect log sheet.

3.Prepare the cleaning solution by adding the amount of cleaner as specified on the labeling to two gallons of pretreated or RO product water.