Millenium™ RO System by GE Osmonics
output after the dwell time.
6.Place the waste hose back in the cleaning solution container.
7.Using the disinfect switch, turn the RO on and operate it for five minutes.
8.Remove the product and waste hoses from the cleaning solution tank and place them in the drain.
4.3.1.3
1.Using the disinfect switch, turn the RO on and run cleaning solution to the drain until the container is empty. This aids in emptying the container of cleaning solution.
CAUTION:
DO NOT ALLOW THE PUMP TO RUN DRY. As an alternative, the solution may be poured in the drain.
2.Rinse and drain the cleaning solution tank and the hoses then let them air dry.
3.Disconnect the disinfect hose from the RO.
4.Ensure that the water supply to the RO unit is on. If it is partially closed, rotate the operate/flush valve to the full flush position.
5.Place the inlet water valve switch in the on position. Let water flow to the drain for one minute.
6.Rotate the operate/flush valve to the operate position.
7.Set the RO pump power switch in the ON position. Rinse for 20 minutes with the Operate/Flush valve in the Operate position until the % rejection and TDS monitors’ displays normalize (e.g. 'usual' % rejection and 'usual' TDS reading). Refer to your operating log for 'usual' readings and compare to the
A.During this rinse period, rotate the Operate/Flush valve to the Flush position briefly and then back to Operate three times to
8.Test the product water for
9.When all of the parameters (% Rejection, TDS, pH) have normalized, operate the RO for an additional 20 minutes in rinse mode.
10.Repeat the cleaning procedure with the other pH cleaner. GE Osmonics recommends using a low pH cleaner first, followed by a high pH cleaner after rinsing thoroughly.