Millenium™ RO System by GE Osmonics
4.3.2BIOSAN™
BIOSAN™ is a liquid cleaner/descaler agent intended for use in removing mineral scale and organic biofilm and is a
CAUTION:
BIOSAN™ is not intended for use in Hemodialysis Machines.
WARNING:
The solution is a mild irritant to eyes and skin. It is recommended that gloves (latex/vinyl) and eye protection be worn when handling BIOSAN™. Neither respiratory protection nor special ventilation are required.
Materials Required:
•BIOSAN™ cleaner
•product water
•clean wide mouth container with volume of 2 ½ gallon or more
•pH test strips
•product rinse line (clear hose with male QD fitting on one end, open on the other)
•“DO NOT USE” warning labels (not provided by Osmonics)
•cleaning / disinfect log sheets
•most current log sheets for water quality comparison
4.3.2.1BIOSAN™ Cleaning Procedure
Read and understand all BIOSAN™ cleaning instructions before beginning procedure.
1.Connect the water inlet line to the water supply, and the waste and product lines to drain. Turn the RO unit and water supply on.
2.Establish the
3.Prepare the BIOSAN™ solution in accordance with the manufacturers instructions.
4.Turn the RO pump power and inlet water valve switches off.
WARNING:
Label machine with appropriate warning signs such as "Do Not Use / Contains BIOSAN™" to prevent use of RO until properly rinsed.
CAUTION:
Water to be used must be chlorine free or damage to the RO membranes may result.
14693e - 31 Mar 03 | 4- 15 | Routine Maintenance |