Troubleshooting
Error indicators
| Indication | Cuff pressure reaches 330mmHg |
|
|
|
| Cause | Twisted bladder inside cuff |
|
|
|
| Correction | Ensure that bladder lies flat within cuff |
|
|
|
| Indication | "Err" displays, cuff pressure releases, then "0 pressure" displays |
|
|
|
| Cause | Excessive arm movement or talking during measurement |
|
|
|
| Correction | Turn monitor OFF then ON and measure again while remaining still |
|
|
|
| Indication | Could not obtain pulse rate |
|
|
|
| Cause | Cuff was too loosely applied, extremely weak pulse, or arrhythmia |
|
|
|
| Correction | Reapply cuff and measure again; if problem persists after cuff is |
beverages, before a measurement.
4.Remove any clothing on the upper arm to place the cuff directly on the skin. Constriction of the upper arm by clothing or a
5.Sit down in a chair with back support and place both feet on the floor.
6.Extend the arm from which the blood
pressure will be taken to the front or side. Rest your hand,
excessive
constriction
correctly positioned, seek medical attention immediately |
Batteries are exhausted. Replace with four new AAA alkaline batteries
Labtron Digital Blood Pressure Monitor | 14 |
Note: Place the cuff at the same level as your heart to obtain accurate measurements. Your heart is located slightly to the right of your left armpit.