Haier HYC-610 Procedure to Test the Alarm System, Trouble Shooting Guide, Solution, Operation

Models: HYC-610

1 10
Download 10 pages 32.8 Kb
Page 10
Image 10
Alarm, Safety, and Auto-diagnostic Features

Operation

Alarm, Safety, and Auto-diagnostic Features

This unit is equipped with all features listed in Table 2. Additionally, it also has auto-diagnostic features.

Table 2, Alarm and Safety Features

Alarm or Safety Refrigerator Behaviors

Alarm Methods and

Buzzer Alarm

 

 

Indication

 

 

 

High

Either upper sensor or

Alarm light flashes

 

Temperature

lower sensor reaches

Buzzer alarm sounds

 

10 and warmer

 

 

 

 

 

 

Low

Either upper sensor or

 

 

lower sensor is below

Alarm light flashes

Buzzer alarm sounds

Temperature

2

 

 

 

 

 

 

 

 

 

 

 

Temperature display

After the time of

 

The unit lost power or

shows for 60 seconds,losing power exceeds

Loss of power

and stops showing

72 hours, buzzer

the power switch is off.

 

 

for 120 seconds. The alarm sounds.

 

 

process repeats until

 

 

 

power is on again.

 

 

 

 

 

Door ajar

Outer door is slightly

After a 10-minute

After a 10-minute

delay, alarm light

delay, buzzer alarm

open or wide open.

 

flashes.

sounds.

 

 

 

 

 

 

 

Upper temperature

Alarm light flashes.

 

 

Temperature display

 

 

sensor circuit shorts

Buzzer alarm sounds.

 

or opens.

shows “E1”.

 

 

 

 

 

 

 

 

 

Lower temperature

Alarm light flashes.

 

 

sensor circuit shorts

Temperature

Buzzer alarm sounds.

 

display shows

 

 

or opens.

 

 

“E2”.

 

 

 

 

Abnormal

 

Alarm light flashes.

 

sensor

Defrost sensor circuit

 

Temperature

Buzzer alarm sounds.

performance

shorts or opens.

display shows

 

 

 

 

 

 

“E3”.

 

 

 

 

 

 

Over temperature

Alarm light flashes.

Buzzer alarm sounds.

 

sensor circuit shorts or

Temperature display

 

opens.

shows “E4”.

 

Rechargeable battery Alarm light stays has low voltage, or on. Display shows

there is no circuit . temperature normally.

Operation

Note: Remote alarm contact and buzzer alarm work together. When the refrigerator is in alarm mode, the remote alarm contact can be disconnected by pushing the buzzer button. If the unit is in the alarm mode due to loss of power, the remote alarm contact remains in the alarm mode. This condition persists until the power is resumed. All alarms will automatically stop once the problem is corrected. If the alarm condition is not corrected within 30 minutes, the buzzer alarm and remote alarm contact are activated again. In the case of losing power, a fully charged battery can support the alarm system for 72 hours. For a newly started refrigerator or a refrigerator that has been unplugged for a long time, you should make sure the battery is fully charged. This charging process can take up to two days. When the battery discharges, the remote alarm contact continues to function.

Procedure to Test the Alarm System

Push the “Alarm Test” button once. The alarm buzzer sounds three consecutive times in 1 Hz frequency. At the same time, the alarm light also flashes three consecutive times and the remote alarm contact opens and closes for three times. The alarm system functions properly.

This blood bank refrigerator should be cared for by dedicated personnel. The operational status of the unit should be inspected and recorded daily. If the temperature runs toward over-temperature condition, you should transfer the stored products to another refrigerator. The products are only allowed back in the original unit only when operating errors are corrected.

Trouble Shooting Guide

Problem

Solution

 

 

 

1.Inspect fuse blocks.

 

 

Unit stops operating.

2.Check that the cord is securely plugged into the power outlet.

 

3.Inspect for problems associated with control wires.

 

 

 

 

4.Make sure the voltage is not too low.

 

 

 

1.Inspect to see if the unit is overloaded or products with warm temperature

 

have been loaded into the unit.

 

2.Make sure there is an air gap between products.

Unit's refrigeration is

3.Check to make sure the unit is not in direct sunlight or subject to any heat

radiation.

not effective.

4.Investigate to see if there have been frequent door openings.

Temperature tends to

operate out of range.

5.Investigate to see if the environment temperature is too high.

 

 

 

6.Inspect air channels to see if they are blocked.

 

 

 

7.Check the condenser air filter to see if it is clogged with dirt.

 

 

 

1.Make sure the unit is placed on a level floor.

The unit is too noisy.

 

2.Make sure any part of the unit does not touch a wall or any object.

 

 

 

 

1.If warm blood products are just loaded to the unit, allow ample time for the

 

temperature to recover. The alarm signal cancels when the temperature

Alarm light flashes,

recovers to normal level.

alarm buzzer sounds.

2.Inspect the door to make sure it is shut properly. Door ajar alarm can sound

 

if the door is slightly opened.

 

3.There might be insufficient power supply. Let it run for some time to recover.

 

 

 

4.Check the temperature of the unit to see if it is in over-temperature condition.

Page 10
Image 10