OPERATION INSTRUCTIONS

02TARE WEIGHT

NOTE: Due to the scale’s sensitivity, we recommend using the REWEIGH function prior to setting the TARE weight, in order to eliminate any operator interference with the item to be tared out.

There are two ways to manually enter a tare weight (such as the weight of a wheelchair, shoes, etc): press MENU (1) and select option 02-Tare Weight or press TARE (9). If there is no weight on the scale platform (value displayed is zero and there is no tare) and the user presses the TARE button (9), the Tare Weight window appears and prompts the user to enter the TARE value and to press ENTER. NOTE: The tare weight must be entered using the following increments: 4oz, 0.2lb, ¼lb. The value entered will display in minus (-). After the TARE has been entered, the scale goes back to normal operation. This TARE value is stored in memory until changed or cleared.

Important: The TARE weight cannot exceed 125 lb.

The number that is to be changed will flash and will move from the left to the right after the appropriate number was entered or by using the c key (left) and the d key (right).

TARE display indicates the weight that has been tared out

Figure 8. Tare Weight

Automatic Tare

The user can set a tare weight by pressing the TARE button (9) while there is a weight on the scale platform. The display will come to zero and TARE will be displayed to indicate there is a tare value in the memory (as displayed in Figure 8 above).

Removing the Tare

Additional pressing of the TARE button will delete the tare value from memory, TARE will disappear from the display and the scale will resume normal operation.

03DATA MANAGEMENT

The scale manages patient data including patient ID, weight, height, tare and BMI. The value is stored in memory or transferred to PC. This function is performed by opening a new data file.

OPENING A NEW DATA FILE:

1.Press the ID button (7).

2.Using the keypad, type in the identification number.

3.Press ENTER.

This scale will also calculate Body Mass Index (BMI).

CALCULATING BMI:

1.Place the patient on the scale platform.

2.Press the BMI button (2).