OPERATOR ACTION

DISPLAY

Turn the scale on. While “START” is displayed on the LCD (for 3 seconds)

“CAL” will appear on the LCD

press the ZERO-KG/LB-KG/LB-ZERO keys one at a time on the keypad.

 

 

 

Press “ENTER”

Enter load weight “Load/200.0”

 

 

Press “ENTER” to calibrate with 200 lb. This can be changed if desired.

“CLEAR”

To change the calibration weight, press the BMI key. Then, use the arrow keys to change the value of the flashing digit.

 

Use the BMI key to advance to the next digit, and again use the arrow keys to change the digit to the desired value.

 

Repeat until you have entered the desired calibration weight. Press “ENTER”

 

The calibration weight must be no less than 100 lb (45 kg) and no more than 300 lb (135 kg).

 

 

 

Clear the weight from the platform and press “ENTER”

Zero Calibration “------------”

Wait without touching or shaking the scale until the zero calibration

Weight calibration “Put XXX.X”

process is finished

 

Load the platform with the required weight and press “ENTER”

Weight calibration “------------”

 

 

Wait without touching or shaking the scale until the zero calibration process is finished.

Calibration “Factor X.XXXX”

Note: If the calibration is done in lb, the factor value you should have is

 

0.9000 – 1.000 If the calibration is done in kg, the factor value should be

 

0.4000 – 0.4250. If the factor value is different repeat the calibration procedure.

 

 

 

Press ENTER

“DONE”/“CLEAR”/XXX

Remove weight from the platform

Rebooting, please wait

Wait until the scale resumes normal operation

“0.0”

 

 

Exploded View of Scale

6

Page 6
Image 6
Health O Meter 752KL, 599KL, 597KL operation manual Exploded View of Scale, Operator Action Display, Done/Clear/Xxx