POTENTIAL FOR ELECTROMAGNETIC INTERFERENCE To avoid inaccurate results caused by electromagnetic interference between electrical and electronic equipments, do not use the device near a cell phone or microwave oven.

23

BPW-040_A.indd 25-26

TROUBLESHOOTING

If any abnormality arises during use, please check the following points:

SYMPTOMS

POSSIBLE

CORRECTION

CAUSES

 

 

 

 

 

 

Batteries have

Replace them

Unit does not

with two new AAA

run down.

alkaline batteries.

turn on when the

 

 

 

START/STOP

Battery polarities

Re-insert the

button is pushed.

have been positioned

batteries in the

 

 

incorrectly.

correct positions.

 

 

 

EE measurement

Cuff has been placed

Wrap the cuff

incorrectly.

properly so that it is

error symbol shown

positioned correctly.

on display or the

 

Did you talk or move

 

blood pressure

 

during

Measure again.

value is displayed

measurement?

Keep wrist steady

excessively low

Shaking of the wrist

during measurement.

(or high).

 

with the cuff on.

 

E1 error symbol

Air circuit

Measure again.

shown on display

abnormality.

 

 

 

 

E2 error symbol

Inflation pressure

Switch the unit off,

shown on display

exceeding 300 mmHg.

then measure again.

 

 

 

E3 error symbol

Error determining

Measure again.

shown on display

measurement data.

 

 

 

 

Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance should you disassemble or attempt to repair the unit by yourself. Contact information for HoMedics Consumer Relations Department can be found on the warranty page (See page 26).