![1 System Description continued](/images/new-backgrounds/51261/5126111x1.webp)
1 System Description continued
System Specifications
Primary Power: Alkaline C cell battery, 48 hours
operation. Note: An audible alarm single beep every 60 seconds indicates that either the Alkaline C cell battery is low or the internal rechargeable NiCad battery is low.
Standby Power: Internal rechargeable NiCad
battery, 6 hours operation. Note: An audible alarm single beep every 60 seconds indicates that either the Alkaline C cell battery is low or the internal rechargeable NiCad battery is low.
Charger: 120 VAC
Display: Backlit LCD
Pump: Miniature internal diaphragm type, flowrate 0.5 SCFH
Alarms: Two concentration alarms, zero drift caution, Low flow and Low battery alarms. Alarms displayed on LCD and indicated by audible beeper.
Standard Accessories: Storage case, 10" extension wand, battery charger, filter material, datalogging software,
Optional Accessories: Analog output cable. Sensor Keeper for storage of additional
Sensor Type: Target gas selective electrochemical
Accuracy: Sensor dependant but generally ± 5% of reading
Repeatability: ± 1% full scale
Operating Ambient Temperature Range: 0o to
+120o F
Storage Temperature: Recommend storing on the charger in an air conditioned office
Weight: 3 lbs. (shipping weight), 1.8 lbs. (detector only)
Dimensions: Approx 3.5" x 5" x 8.5"
Outputs:
Memory: 12,000 data points, storage interval between points programmable at 1, 5, 10, or 15 minutes. Typical capacity: 8 days at 1 minute storage interval
Manning | 6 |