System Specifications

Primary Power: Alkaline C cell battery, 48 hours

operation. Note: An audible alarm single beep every 60 seconds indicates that either the Alkaline C cell battery is low or the internal rechargeable NiCad battery is low.

Standby Power: Internal rechargeable NiCad

battery, 6 hours operation. Note: An audible alarm single beep every 60 seconds indicates that either the Alkaline C cell battery is low or the internal rechargeable NiCad battery is low.

Charger: 120 VAC plug-in charger for internal standby battery

Display: Backlit LCD

Pump: Miniature internal diaphragm type, flowrate 0.5 SCFH

Alarms: Two concentration alarms, zero drift caution, Low flow and Low battery alarms. Alarms displayed on LCD and indicated by audible beeper.

Standard Accessories: Storage case, 10" extension wand, battery charger, filter material, datalogging software, RS-232 output cable (DB9 connector)

Optional Accessories: Analog output cable. Sensor Keeper for storage of additional SMART-CELLs

Sensor Type: Target gas selective electrochemical

SMART-CELL

Accuracy: Sensor dependant but generally ± 5% of reading

Repeatability: ± 1% full scale

Operating Ambient Temperature Range: 0o to

+120o F

Storage Temperature: Recommend storing on the charger in an air conditioned office

Weight: 3 lbs. (shipping weight), 1.8 lbs. (detector only)

Dimensions: Approx 3.5" x 5" x 8.5"

Outputs: RS-232 output of stored gas values. Analog output 0 to 1.0 VDC (requires optional analog output cable)

Memory: 12,000 data points, storage interval between points programmable at 1, 5, 10, or 15 minutes. Typical capacity: 8 days at 1 minute storage interval

Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.