49
SCALE OPERATION
Wire Security Seal (Seal Conversion Kit 46-46890)*
1. Temporarily remove the platter and place it in a safe location.
2. Install the calibration switch/button tabbed cover.
3. Secure the cover in place with the flat head Phillips screw provided.
4. Position the security seal cage over the tab on the calibration switch/button cover.
5. Thread the wire through the hole in the tab on the calibration switch/button cover and through the wire lock at the other end
forming a loop (see Figure 67).
6. Crimp the lock closed to secure the wire, then pack the wire and lock into the cage.
7. Insert the tab on the end of the transparent cage cover into the slot on the security cage then snap the other end into
place over the security cage hook.
8. Reinstall the platter.
Figure 67. Calibration Switch/Button Wire Security Seal Assembly.
* These procedures are for MS2320 StratosH Scanner/Avery Scale models only.