Additions to the User Manual

The additions to User Manual 34-ST-33-39 that relate to the new Gauge Pressure transmitter models are given in Table 1 of this addendum. Use the information in Table 1 to reference and annotate your Installation Guide.

Table 1 – Additions to the User Manual

 

 

Page # in User

 

Sub-Section

 

Description of Change

 

 

 

 

Manual

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

18

 

3.1 Considerations for ST

In the left column of Table 5, under the heading

 

 

 

 

 

3000 Transmitter

Gauge Pressure, add the information as

 

 

 

 

 

Temperature Limits

indicated by the highlights in Exhibit A, below.

 

 

 

 

 

(Note: Ranges for Ambient Temperature and

 

 

 

 

 

Table 5

Operating

 

 

 

 

 

Process Interface Temperature are the same as

 

 

 

 

 

Temperature Limits

 

 

 

 

 

for other models in each series.)

 

 

 

 

 

 

(Transmitters with Silicone

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fluid Fill Fluids)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

19

 

3.1 Considerations for ST

In the row of Table 6 titled Gauge Pressure, add

 

 

 

 

 

3000 Transmitter

the information as highlighted in Exhibit B, below.

 

 

 

 

 

Pressure Ratings

 

 

 

 

 

 

 

 

 

 

Table 6

Transmitter

 

 

 

 

 

 

 

 

 

 

Overpressure Ratings

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

47

 

5.1 Wiring Diagrams and

In the leftmost column of the table on page 47,

 

 

 

 

 

Dimension Drawings

add the information as hignhlighted in Exhibit C,

 

 

 

 

 

 

 

 

below.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Exhibit A –Additions to Table 5

 

 

 

 

 

 

 

 

 

 

 

Transmitter Type and Model

Ambient Temperature

 

Process Interface Temperature

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

°C

 

°F

 

°C

 

°F

 

 

 

 

 

 

 

 

 

 

 

 

 

Draft Range

STD110

-40 to 70

 

-40 to 158

 

-40 to 70

 

-40 to 158

 

 

Differential PressureSTD125

-40 to 85

 

-40 to 185

 

-40 to 85

 

-40 to 185

 

 

STD120, STD130, STD170

-40 to 93

 

-40 to 200

 

-40 to 125

 

-40 to 257

 

 

 

STD904, STD924,

 

 

 

 

 

 

 

 

 

 

STD930, STD974

-40 to 85

 

-40 to 185

 

-40 to 125

 

-40 to 257

 

 

Gauge Pressure

 

 

 

 

 

 

 

 

 

 

 

STG140, STG170, STG180,

-40 to 93

 

-40 to 200

 

-40 to 125

 

-40 to 257

 

 

STG14L, STG17L, STG18L,

 

 

 

 

 

STG19L

STG14T

-40 to 93

 

-40 to 200

 

-40 to 150 †

 

-40 to 302 †

 

 

 

 

-15 to 65

 

5 to 149

 

-15 to 95 ††

 

5 to 203 ††

 

 

 

 

STG93P

-40 to 85

 

-40 to 185

 

-40 to 125

 

-40 to 257

 

 

 

STG944, STG974

 

 

 

 

 

 

 

 

 

 

STG90L, STG94L,

-40 to 85

 

-40 to 185

 

-40 to 110

 

-40 to 230

 

 

STG97L, STG98L, STG99L

 

 

 

 

 

 

 

 

 

Absolute Pressure

STA122

-40 to 93

 

-40 to 200

 

See Specification Sheet

 

 

 

 

STA140

-40 to 93

 

-40 to 200

 

-40 to 80

 

-40 to 176

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4/02

 

 

34-ST-99-15 Addendum

 

 

 

2 of 4