Exhibit B –Additions to Table 6

Transmitter Type

Upper Range Limit

Maximum Working

Overpressure Rating

 

(URL)

Pressure Rating

 

 

 

 

 

Draft Range

10 inches H2O (25

50 psi (3.5 bar)

50 psi (3.5 bar)

 

mbar)

 

(No overpressure

 

 

 

protection is provided)

 

 

 

 

Differential Pressure

400 inches H2O (1 bar)

3000 psi (210 bar)

3000 psi (210 bar)

 

100 psi (7 bar)

3000 psi (210 bar)

3000 psi (210 bar)

 

 

 

 

 

3000 psi (210 bar)

3000 psi (210 bar)

3000 psi (210 bar)

 

 

 

 

Gauge Pressure

100 psi (7 bar)

100 psi (7 bar)

150 psi (10.3 bar)

 

 

 

 

 

300 psi (21 bar)

300 psi (21 bar)

450 psi (31 bar)

 

 

 

 

 

500 psi (35 bar)

500 psi (35 bar)

750 psi (52 bar)

 

 

 

 

 

3000 psi (210 bar)

3000 psi (210 bar)

4500 psi (310 bar)

 

 

 

 

 

6000 psi (415 bar)

6000 psi (415 bar)

9000 psi (620 bar)

 

 

 

 

 

10000 psi (690 bar)

10000 psi (690 bar)

15000 psi (1034 bar)

 

 

 

 

Absolute Pressure

780 mmHg Absolute

780 mmHg Absolute

Full vacuum to 1550

 

(1 bar)

(1 bar)

mmHg Absolute

 

 

 

(2 bar)

 

 

 

 

 

500 psia (35 bar)

500 psia (35 bar)

750 psia (52 bar)

 

 

 

 

3 of 4

34-ST-99-15 Addendum

4/02