3.2 Considerations for SFC/SCT, Continued

Install SFC battery pack, continued

Temperature Limits

Usage guidelines

 

Table 7

 

Installing and Charging SFC Battery Pack, Continued

 

 

Step

 

 

Action

 

 

 

 

 

5

Plug battery charger into any standard 120 Vac outlet or universal-

 

 

 

European 240 Vac outlet as applicable for charger power rating. If

 

 

 

240 Vac charger is supplied with stripped leads instead of universal-

 

 

 

European plug, lead identification for 240 Vac charger is as follows.

 

 

 

 

 

 

 

 

 

 

 

Lead Color…

 

Function…

 

 

 

 

Blue

 

Neutral

 

 

 

 

Brown

 

Hot

 

 

 

 

Green/Yellow

 

Ground

 

 

 

 

 

 

 

 

 

 

 

ATTENTION

It takes up to 16 hours to fully recharge the battery

 

 

 

pack and you can use the SFC continuously for up to 24 hours

 

 

 

before the battery pack needs recharging.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The ambient operating temperature limits for the SFC are -10 to 50 °C (14 to 122 °F) with relative humidity in the range of 10 to 90% RH.

For transmitters operating in the Analog mode, be sure to put an analog control loop into its manual mode before initiating SFC communications with the transmitter. Also, be sure any switches that may trip alarms or interlocks associated with the analog loop are secured or turned OFF. Communication superimposes digital signals on the loop wiring that could affect the analog control signal.

Be sure the power supply voltage does not exceed 45Vdc. The ST 3000 transmitter and SFC were designed to operate with voltages below 45Vdc.

Be sure there is at least 250 ohms of resistance between the SFC and the power supply for proper communications.

SCT 3000 Requirements

The Smartline Configuration Toolkit (SCT 3000) consists of the software program which is contained on diskettes and a Smartline Option Module which is the hardware interface used for connecting the host computer to the ST 3000 transmitter.

Be certain that the host computer is loaded with the proper operating system necessary to run the SCT program. See the SCT 3000 Smartline Configuration Toolkit Start-up and Installation Manual 34-ST-10-08for complete details on the host computer specifications and requirements for using the SCT 3000.

2/05

ST 3000 Release 300 Installation Guide

21