3.2Considerations for SFC/SCT

Install SFC battery pack

If the SFC battery pack was removed for shipping and/or storage, you will have to install the battery pack and charge the batteries before you can operate the SFC.

The procedure in Table 7 outlines the steps for the battery pack.

Table 7 Installing and Charging SFC Battery Pack

Step

Action

1Turn SFC face down on working surface. Use metric hex wrench (2.5 mm) to remove screws in battery compartment cover and remove cover.

2Insert battery pack in compartment and connect plug in compartment to pin on battery back

Example - Battery pack installation.

Battery Pack

 

Hex Screws

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

21004

3Replace cover and tighten hex screws

4Connect lead from battery charger to recessed connector on left side of SFC.

WARNING The SFC battery charger is not intrinsically safe. Always recharge the SFC battery pack in a nonhazardous location. The SFC itself is an intrinsically safe device.

Continued on next page

20

ST 3000 Release 300 Installation Guide

2/05