3.1Considerations for ST 3000 Transmitter, Continued
Temperature limits | Table 5 lists the operating temperature limits for the various types of | ||||||
|
| transmitters with silicone fill fluids. See transmitter specifications for | |||||
|
| temperature limits of transmitter with alternative fill fluids. |
| ||||
Table 5 | Operating Temperature Limits (Transmitters with Silicone Fill Fluids) |
| |||||
|
|
|
|
| |||
Transmitter Type and Model | Ambient Temperature | Process Interface Temperature | |||||
|
|
| °C | °F | °C |
| °F |
Draft Range | STD110 |
| |||||
Differential PressureSTD125 |
| ||||||
STD120, STD130, STD170 |
| ||||||
| STD904, STD924, |
|
|
|
|
| |
| STD930, STD974 |
| |||||
Gauge Pressure |
|
|
|
|
|
| |
STG140, STG170, STG180, |
|
|
|
|
| ||
STG14L, STG17L, STG18L |
| ||||||
|
| STG14T |
| ||||
|
| STG93P | 5 to 149 |
| 5 to 203 †† | ||
| STG944, STG974 |
| |||||
| STG90L, STG94L, |
|
|
|
|
| |
| STG97L, STG98L |
| |||||
Absolute Pressure | STA122 | See Specification Sheet | |||||
|
| STA140 |
| ||||
|
| STA922 | See Specification Sheet | ||||
|
| STA940 |
| ||||
Flange Mounted |
|
|
|
|
|
| |
STF128, STF132, STF924, |
| ||||||
STF932 |
|
|
| ||||
|
|
|
|
| |||
STF12F, STF13F, STF92F, |
| ||||||
STF93F |
|
|
| ||||
STF14F |
|
|
| ||||
Gauge Pressure Flange Mount |
| ||||||
STF14T |
|
|
| ||||
Remote Diaphragm Seals |
|
|
|
|
| ||
STR12D, STR13D, STR14G, |
|
|
|
|
| ||
| STR17G, STR14A | See Specification Sheet | See Specification Sheet | ||||
| STR93D, STR94G | See Specification Sheet | |||||
†Process temperatures above 125 °C (257 °F) require a reduction in the maximum ambient temperature as follows:
Process Temperature | Ambient Temperature Limit |
150 °C (302 °F) | 50 °C (122 °F) |
140 °C (284 °F) | 60 °C (140 °F) |
125 °C (257 °F) | 85 °C (185 °F) |
††Process temperatures above 65 °C (149 °F) require a 1:1 reduction in maximum ambient temperature. NOTE: For transmitters with local meter option see Table
NOTE: Transmitters with other fill fluids (CTFE, Neobee, Etc.) have different Operating Temperature Limits. For more specific information, refer to the appropriate Specification and Model Selection Guide or transmitter nameplate
18 | ST 3000 Release 300 Installation Guide | 2/05 |