2.6Setting Range Values Using Local Adjustments

Local zero and span option

About local adjustments

For transmitter applications that do not require an SFC nor digital integration with Honeywell’s TPS systems, ST 3000 transmitters are available with optional local zero and span adjustments.

The transmitter’s range values can be set by using the pushbuttons on the face of the local zero and span option or smart meter. Refer to the procedure for setting the range values to applied pressures using local zero and span adjustments in Appendix A Smart Meter Reference in this guide.

16

ST 3000 Release 300 Installation Guide

2/05