2.4Changing Default Failsafe Direction, Continued
Procedure, continued
Table 4 | Cutting Failsafe Direction Jumper | |
|
|
|
| Step | Action |
|
|
|
| 1 | With transmitter on bench and no power applied. Loosen |
|
| lock and unscrew end cap from electronics side of transmitter |
|
| housing. |
|
|
|
| 2 | • If applicable, unsnap Local Smart Meter from PWA mounting |
|
| bracket and unplug cable from connector on back of meter |
|
| assembly. |
|
| • Loosen two retaining screws and carefully pull mounting bracket |
|
| and PWA from housing. |
|
| • Using retaining clip remove |
|
| • Remove |
|
| PWA and mounting bracket assembly. |
|
|
|
| 3 | With component side of PWA facing you, locate failsafe jumper W1 |
|
| and cut it in half with small wire cutter such as dykes. See Figure 2. |
|
| This changes failsafe action from upscale to downscale. |
|
|
|
| 4 | Reverse applicable previous steps to replace PWA. |
|
|
|
| 6 | Turn ON transmitter power. |
|
|
|
Figure 2 Location of Failsafe Direction Jumper on PWA.
Power
Connector



Meter
Connector
Failsafe
Direction
Jumper 

14 | ST 3000 Release 300 Installation Guide | 2/05 |