2.5Optional Write Protect Jumper

Write protect option

The ST 3000 transmitters are available with a “write protect option”. It

 

consists of a jumper located on the transmitter’s PWA that you can

 

position to allow read and write access or read only access to the

 

transmitter’s configuration database. When the jumper is in the read

 

only position, you can only read/view the transmitter’s configuration

 

and calibration data. Note that the factory default jumper position is for

 

read and write access.

 

There is no need to check the jumper position unless you want to change

 

it. Refer to the steps in Table 4 to remove the PWA from the transmitter

 

and Figure 3 to reposition the jumper.

 

Figure 3 shows the location of the write protect jumper on the PWA for

 

Release 300 transmitters.

Figure 3 Write Protect Jumper Location and Selections.

W

 

 

 

Read and

 

R

 

 

 

Write

 

 

 

 

 

 

 

Read Only

Power

Connector

Meter

Connector

Failsafe

Direction

Jumper