HP 8673 manual Performance Tests

Models: 8673

1 725
Download 725 pages 22.83 Kb
Page 30
Image 30

HP 8673C/D

Performance Tests

SECTION IV, Part 2

PERFORMANCE TESTS

4-I. INTRODUCTION

In order to provide Volume 2 as soon as possible, the Full Performance Tests are omitted from this preliminary Service Manual.

Section IV, Part 1,contains the Abbreviated Per- formance Tests. These tests should be performed after repairing the Signal Generator. They can also be used for incoming inspection and preven- tative maintenance. They are not intended to be a

complete check of specifications, but they will provide a limited calibration.

NOTE

Specifications and characteristics apply after a 1 hour warm-up, ouer the tempera- ture range of0 to 55” C (exceptspecifica- tions for harmonically related spurious signals, RFoutput leuel,pulsepeak level accuracy, and amplitude modulation; which apply over the ran@ 41.5 to +35” C), after an AUTO PEAK opera tion has beenperformed. For additional

information concerning the use of AUTO PEAK, refer to Sections 3-11and 3-12.

Page 30
Image 30
HP 8673 manual Performance Tests