Page 8
Required Equipment
Required equipment is listed in Table B. If the recommended
models are not available, equipment with equivalent specifications may be used. Repairs or servicing should be performed only by qualified personnel.
Table B. Required Equipment
Equipment | Required Characteristics | Recommended | |
Model | |||
|
| ||
|
|
| |
Calibrator | AC Voltage Range: | Fluke 5500 or | |
| Accuracy: ±0.07% (Basic) | Wavetek 9100 | |
| Frequency Range: 40 ~ 100KHz | Calibrator or | |
| Accuracy: ± 2% | equivalent | |
| DC Voltage Range: |
| |
| Accuracy: ±0.006% (Basic) |
| |
| Current Range: 0 ~ 10A |
| |
| Accuracy: |
| |
| AC (40Hz to 1KHz): ±0.08% (Basic) |
| |
| DC: ± 0.02% (Basic) |
| |
| Frequency Source: |
| |
| 5.00Hz ~ 1.0000MHz |
| |
| Accuracy: ±0.001% |
| |
| Amplitude: |
| |
| 0.5V |
| |
| Accuracy: ± 5% |
| |
| Ω range: 1Ω ~ 2G |
| |
| Accuracy: ±0.03% (Basic) |
| |
| Capacitance Range: 1PF ~ 40mF |
| |
| Accuracy: ±0.10% (Basic) |
| |
| Temperature Range: |
| |
| Accuracy: ±0.3 (Basic) |
|
Form Number | Rev 5 Dec 2006 |