Page 21

ACV High Frequency Calibration

Calibrating the ACV Frequency Response

1.400V Range

Set the calibrator to generate 200.00V / 10KHz ±0.07% at least, adjust VC3 to have a reading of 200.50± 20 count .

2.40V Range

Set the calibrator to generate 20.000V /10KHz±0.07% at least, adjust VC2 to have a reading of 20.050± 20 count .

3.4V Range

Set the calibrator to generate 2.0000V /10KHz±0.07% at least, adjust VC1 to have a reading of 2.0050± 20 count .

Form Number TM61633-5

Rev 5 Dec 2006