Page 17
Table 9. AC mA Test |
|
|
|
| ||
|
|
| Input |
| Readings | |
| Step |
| Current | Frequency | ||
| 1 |
| 36.000mA | 50 Hz | 35.488 to 36.512 | 35.632 to 36.368 |
| 2 |
| 36.000mA | 400 Hz | 35.488 to 36.512 | 35.632 to 36.368 |
| 3 |
| 360.00mA | 50 Hz | 354.88 to 365.12 | 356.32 to 363.68 |
| 4 |
| 360.00mA | 400Hzz | 354.88 to 365.12 | 356.32 to 363.68 |
Testing the Amp (A) Function
To verify the accuracy in the ampere (A) measurement function, do the
Following:
1.Connect the Calibrator to the A and COM inputs of the Meter.
2.Turn the rotary switch to A
3.Apply the inputs for steps
4.For each input, compare the readings on the Meter display to the display readings in Table 10.
5.If the display reading falls outside of the range shown in Table 10, the meter does not meet specification.
Table 10. DCA Test |
|
| ||
| Input |
| Readings |
|
| Step | Current | ||
| 1 | 3.6000A | 3.5780 to 3.6220 | 3.5888 to 3.6112 |
| 2 | 9.000A | 8.9150 to 9.0850 | 8.9420 to 9.0580 |
6.Press the blue button on the Meter to toggle to the AC amp measurement function.
7.Apply the inputs for steps
8.For each input, compare the reading on the Meter display to the display readings in Table 11.
9.If the display reading falls outside of the range shown in Table 11, the meter does not meet specification.
Table 11. ACA Test |
|
|
| |||
|
| Input |
| Readings | ||
| Step | Current |
| Frequency | ||
| 1 | 3.6000A |
| 50 Hz | 3.5488 to 3.6512 | 3.5632 to 3.6368 |
| 2 | 3.6000A |
| 400 Hz | 3.5488 to 3.6512 | 3.5632 to 3.6368 |
| 3 | 9.000A |
| 50 Hz | 8.8120 to 9.1888 | 8.8480 to 9.1520 |
| 4 | 9.000A |
| 400 Hz | 8.8120 to 9.1888 | 8.8480 to 9.1520 |
Testing the Frequency Function
To verify the accuracy of the Meter’s frequency function, do the following: 1. Connect the Calibrator to the VΩ and COM inputs on the Meter.
Note: The accuracy of the Calibrator’s frequency function must be appropriate for the specified accuracy of the Meter.
2.Set the rotary switch to Hz.
3.Set the Function Generator for the square wave voltage and frequency for steps
4.Compare the reading on the Meter display with the display reading shown in Table 12.
5If the display reading falls outside of the range shown in Table 12, the Meter does not meet specification.