Testing the Capacitance Function (cont’d)

d)For each input, compare the readings on the Meter display to display readings in Table 7. The meter selects the proper range automatically. Each measurement takes about one second per range. 5mF will taken about 15 seconds.

3.If the display reading falls outside of the range shown in Table 7, the Meter does not meet specification.

Table 7. Capacitance Test

 

 

 

 

 

Input

 

Readings

 

Step

Capacitance

61-633

 

61-635

 

1

3.600nF

3.512 to 3.688

 

3.548 to 3.652

 

2

36.00nF

35.12 to 36.88

 

35.48 to 36.52

 

3

360.0nF

351.2 to 368.8

 

354.8 to 365.2

 

4

3600μF

3.512 to 3.688

 

3.548 to 3.652

 

5

36.00μF

34.76 to 37.24

 

35.12 to 36.88

 

6

360.0μF

347.6 to 372.4

 

351.2 to 368.8

 

7

1.000mF

.9410 to 1.059

 

.9510 to 1.049

Checking the Diode Test Function

To check the diode test function, do the following:

1.Connect the Calibrator to the Vand COM inputs on the Meter.

2.Turn the rotary switch to .

3.Apply 3.000V.

The meter display should read approx. 3.000V dc.

4.Press “Blue” key to test function.

5.Apply a 50resistor to meter, the built-in beeper alarms.

Testing the Milliamp (mA) Function

To verify the accuracy of DC and AC current measurement functions, do

The following:

1.Connect the Calibrator to the mA and COM inputs on the Meter.

2.Turn the rotary switch to mA.

3.Apply the inputs for steps 1-2 in Table 8.

4.For each input, compare the readings on the Meter display to the display readings in Table 8.

Table 8. DC mA Test

 

 

 

 

Input

Readings

 

Step

Current

61-633

61-635

 

1

36.000rnA

35.780 to 36.220

35.888 to 36.112

 

2

360.00mA

357.80 to 362.20

358.88 to 361.12

5.Press the blue button on the Meter to toggle to AC measurement function.

6.Apply the inputs for steps 1-4 in Table 9.

7.For each input, compare the readings on the Meter display to the display readings in Table 9.

8.If the display reading falls outside of the range shown in the Tables, the meter does not meet specification.

Form Number TM61633-5

Rev 5 Dec 2006