ADVANCED FEATURES
USING THE DIAGNOSTIC FEATURE
Reset. With the many User Options available to customize the 350TX, it is easy to configure the unit in such a way that it is detrimental to a particular use. By using the Reset function, all variable or user- controlled features (except Filter) of the 350TX are returned to the factory settings.
The Reset function is an important first step in
troubleshooting problems to ensure that a user selected setting is not the cause for the perceived problem.
USING THE DIAGNOSTIC FEATURE
The 350TX contains a powerful diagnostic tool that can aid in determining the cause of a problem.
To use the Diagnostic feature, select the Diagnostic feature on the
Self Test. Initially, the 350TX will perform a
appear. PASSED indicates the internal test discovered no failures, FAILED indicates a significant internal problem was discovered and the unit requires factory service. A code number appears that tells
repair personnel where the problem is located.
Transducer Connection. The first connection tested is Transducer input. If a transducer is detected, the message CONNECTED appears. If a series of dashed lines is shown, no transducer is connected or there is a problem with the transducer or cable.
If you are using multiple transducers, switch between transducers to confirm the connections for each.
25