SECTION 5—INDICATOR LIGHTS
Part No 1100873 32 HomeFill™II Compressor
SECTION 5—INDICATOR LIGHTSINDICATOR LIGHT
COLOR COMPRESSOR
STATUS EFFECT
None Compressor is Off O2 flowing to user if
concentrator is on
O2 BELOW NORMAL -
YELLOW
Compressor is On and
warming up
(0-3 minutes after
power is On)
Cylinder filling not
started
O2 flowing to user if
concentrator is On
O2 from concentrator
is < 9 0% (a ny t ime a fte r
three minute warmup)
Cylinder filling stops,
will restart when O2
> 90%
O2 flowing to user if
concentrator is On
FILLING - GREEN Compressor is On and
cylinder is filling. O2
from concentrator is
>90% (any time after
three minute warmup)
Cylinder is filling
O2 flowing to user if
concentrator is On
FULL - GREEN Compressor is On.
Cylinder is finished
filling (full) -remove
cylinder (FULL light is
On in >10 mins.)
or
Cylinder is not
attached properly -
check connection
(FULL light is On in
<10 mins.)
Cylinder filling stops
Cannot fill another
cylinder until
compressor is turned
Off and On again
O2 flowing to user if
concentrator is On