SECTION 5—INDICATOR LIGHTS

Part No 1100873 32 HomeFillII Compressor
SECTION 5—INDICATOR LIGHTS

INDICATOR LIGHT

COLOR COMPRESSOR

STATUS EFFECT

None Compressor is Off O2 flowing to user if
concentrator is on
O2 BELOW NORMAL -
YELLOW
Compressor is On and
warming up
(0-3 minutes after
power is On)
Cylinder filling not
started
O2 flowing to user if
concentrator is On
O2 from concentrator
is < 9 0% (a ny t ime a fte r
three minute warmup)
Cylinder filling stops,
will restart when O2
> 90%
O2 flowing to user if
concentrator is On
FILLING - GREEN Compressor is On and
cylinder is filling. O2
from concentrator is
>90% (any time after
three minute warmup)
Cylinder is filling
O2 flowing to user if
concentrator is On
FULL - GREEN Compressor is On.
Cylinder is finished
filling (full) -remove
cylinder (FULL light is
On in >10 mins.)
or
Cylinder is not
attached properly -
check connection
(FULL light is On in
<10 mins.)
Cylinder filling stops
Cannot fill another
cylinder until
compressor is turned
Off and On again
O2 flowing to user if
concentrator is On