SECTION
2.Ensure toggle valve is open. Open the toggle valve by rotating counterclockwise.
3.Examine the gauge on the connected conserver or regulator. Perform one of the following:
•If the cylinder pressure is less than 1500 psig, fill the cylinder. Connect the cylinder to the compressor. Refer to Filling the Cylinder on page 13.
•If the cylinder pressure is greater than 1500 psig., DO NOT attempt to fill this cylinder. It may not fill. Use cylinder as normal.
Filling the Cylinder
NOTE: For this procedure, refer to FIGURE 4.2 thru FIGURE 4.7.
1.Check the cylinder pressure before attempting to fill cylinder. Refer to Checking Cylinder Pressure on page 12.
2.Close the toggle valve by rotating clockwise (FIGURE 4.2).
3.Remove the conserver or regulator according to the device manufacturer instructions (FIGURE 4.2).