SECTION 6—MAINTENANCE
Perfecto2
™Series 26 Part No 1143482
FIGURE 6.1 Cleaning the Cabinet Filter
Cabinet
WARNING
UNPLUG the concentrator when cleaning. To avoid
electrical shock, DO NOT remove cabinet.
1. Clean the cabinet with a mild household cleaner and
non-abrasive cloth or sponge.
Humidifier
NOTE: To clean the oxygen humidifier, follow the instructions
provided by the manufacturer. If none are provided, follow these
STEPS:
1. Clean the humidifier every day.
2. Wash it in soapy water and rinse it with a solution of ten
parts water and one part vinegar.
3. Rinse thoroughly with hot water and refill with distilled
water to the level shown on the humidifier.
WARNING
DO NOT overfill humidifier bottle.